Questions And Answers

# List All Questions Search List by Category
Question In calibrating the Combo Tester (model # 41203) we observed that the standard called for in Tech Brief PS-2058, [the calibration box] model# 41230 does not provide a 4 to 1 [accuracy] ratio [which is the industry standard]. The Combo Tester has a rated accuracy of 5%, while the standard has a rated accuracy of 10%. Is this a typo or are these accuracy's correct? - Bob, Nashua, NH
Answer This was a typo. The correct accuracies are: Combo Tester, 10% Calibration Box, 2%. Thanks for your feedback.
Related Categories:
 
If you have found this Q/A useful, please rate it based on its helpfulness.
Rating Rating Rating Rating Rating
This question has been rated: 95%95%95%95%95%95%
(95% at 4 Ratings)